Figure 2.10: set point thickness parameter, Setpoint thickness parameter, Figure 2.11: set point timer parameter – INFICON STM-100/MF Thin Film Deposition Monitor User Manual
Page 27: Setpoint timer
![background image](/manuals/562496/27/background.png)
SETPOINT THICKNESS
Figure 2.10: Set Point Thickness Parameter.
LEGEND
SP THK
RANGE
0 to 9999
UNITS
KILO ANGSTROMS
The SETPOINT THICKNESS parameter is used to provide a comparison
point for the STM-100 / MF Thickness Setpoint Relay. This set of relay
contacts will be closed whenever the thickness display equals or exceeds the
setpoint value. A setpoint value of 0.000 kÅ causes the thickness setpoint
function to be ignored.
SETPOINT TIMER
Figure 2.11: Set Point Timer Parameter.
LEGEND
SP TMR
RANGE
00:00 to 99:59
UNITS
MIN:SEC
The SETPOINT TIMER parameter is used to provide a comparison point
for the STM-100 / MF Timer Setpoint Relay. The internal STM-100 / MF timer
is pre-set with the setpoint value whenever the front panel ZERO key, or a
remote ZERO TIMER input, is activated. The Timer Relay contacts are
opened at this time and the timer display begins counting downward toward
zero. When the timer display reaches zero the relay contacts are closed and
the display timer begins counting in an upwards fashion. A setpoint value of
00:00 will cause the timer setpoint function to be ignored.
Setpoint Thickness
Parameter
Setpoint Timer
Page 2 - 7
OPERATION AND PROGRAMMING