Command list for stm-100 / mf, Table 5.2: communication command summary, Detailed command description – INFICON STM-100/MF Thin Film Deposition Monitor User Manual
Page 63: Section 5.6, Software version code, Shutter relay control, Zero thickness and timer, Computer interfacing page 5 - 8, Parameter example comment
Parameter
Example
Comment
Sent
Returned
@
None
@
STM100C5
Firmware Version and inst. model number
A
[@,!,?]
A@
Turn shutter relay off
B
None
B
Zeros timer and thickness
C
None
C
Zeros thickness
D
None
D
Zeros timer
E
=[0.500 - 99.99],[?]
E=1.23
Set current film density to 1.23
F
=[0.100 - 9.999],[?]
F?
1.234
Returns current film Z factor
G
=[0 - 9999999],[?]
G=550
Set current film end thickness to 550Å
H
=[0 - 9999999],[?]
H=10560
Set current film setpoint thickness to 10.56KÅ
I
=[00:00 - 99:59],[?]
I=15:30
Set current film setpoint timer to 15:30
J
=[10.0 - 399],[?]
J = 80.1
Set current film tooling to 80.1%
K
[@,!,?]
K?
@
Return test mode status, was off
L
None
L
Acknowledge "a" response
M
None
M
!
Return crystal fail status, was failed
O
None
O
!
Return setpoint timer relay, was closed
P
None
P
@
Return end thickness relay, was open
Q
None
Q
C
Return remote inputs, ZERO TIMER,
ZERO THICK active
R
None
R
193
Return config switch, switches 5,6,12 on
S
None
S
-0001595
Return thickness value -1.595KÅ
T
None
T
(sp)012.4
Return rate, 12.4 Å/s
U
None
U
5319234
Return sensor freq., 5,319,234Hz
V
None
V
012.4
Return crystal life, 12.4% left
W
None
W
+12:45
Return timer, 12 min 45 sec counting up
X
None
X
(sp)0000201
Return log thickness, 201 Å
Y
None
Y
+49:21
Return log timer, since THICK ZERO.
Z
None
Z
-012.3
Return log rate, -12.3 Å/sec
a
None
a
A
Return RESET Status, true
b
None
b
Set parameters to default values
c
[@,!,?]
c?
!
Query beeper status, returned active
d
None
Internal use commands
e
Mask
Internal use commands
f
None
Internal use commands
g
None
Internal use commands
h
None
Internal use commands
i
[1-9,?]
i5
Sets the current film to number 5.
j
[1-9],[0.500 - 99.99]|[?]
Sets / queries the selected film density.
k
[1-9],[0.100 - 99.99]|[?]
Sets / queries the selected film z-factor.
l
[1-9],[0 - 9999000]|[?]
Sets / queries the selected film End Thickness
m
[1-9],[0 - 9999000]|[?]
Sets / queries the selected film
set point relay thickness
n
[1-9],[00:00 - 99:59]|[?]
Sets / queries the selected film timer relay
o
[1-9],[10.0 - 399]|[?]
Sets / queris selected film tooling factor
Table 5.2: Communication Command Summary.
DETAILED COMMAND DESCRIPTION
_________________________________________________________________________
Command:
@
Parameter:
None
Description:
Returns a string indicating the product model number and the software revision level.
The form is "STM100XY" where X is major (A.B,..) and Y is minor digits (0..9).
_________________________________________________________________________
Command:
A
Parameter:
[@,!,?]
Description:
Turns the shutter relay on and off. A "@" will turn the shutter off and a "!" will turn it
on. The indicator on the LCD will indicate the new state. A "?" will send back "@" or
"!" as the current status.
_________________________________________________________________________
Command:
B
Parameter:
None
Description:
Zeros the timer and thickness. This duplicates the ZERO button on the front panel.
Command List For
STM-100 / MF
SECTION 5.6
Software Version
Code
Shutter Relay Control
Zero Thickness and
Timer
COMPUTER INTERFACING
Page 5 - 8