Turn the test mode on/off, Acknowledge the status of non-volitile memory, Get the crystal fail status – INFICON STM-100/MF Thin Film Deposition Monitor User Manual
Page 65: Get the status of the setpoint timer relay, Get the status of the setpoint thickness relay, Computer interfacing page 5 - 10
![background image](/manuals/562496/65/background.png)
_________________________________________________________________________
Command:
K
Parameter:
[@,!,?]
Description:
Turns the test mode on and off. A "@" will turn the test mode off, a "!" will turn it on,
and a "?" will return the current status.
_________________________________________________________________________
Command:
L
Parameter:
None
Description:
Acknowledges valid receipt of power on status and resets internal flags. See "a"
command.
_________________________________________________________________________
Command:
M
Parameter:
None
Description:
Returns the crystal fail status. If a "@" is returned, the crystal is good. A status of "!"
indicates that the crystal has failed.
_________________________________________________________________________
Command:
N
Parameter:
None
Description:
Returns the status of the setpoint timer relay. If a "@" is returned, the relay is open
(time has not been reached). A status of "!" indicates that the relay is closed (time
has been reached).
_________________________________________________________________________
Command:
O
Parameter:
None
Description:
Returns the status of the setpoint thickness relay. If a "@" is returned, the relay is
open (thickness has not been reached). A status of "!" indicates that the relay is
closed (time has been reached).
Turn the Test Mode
On/Off
Acknowledge the
Status of Non-Volitile
Memory
Get the Crystal Fail
Status
Get the Status of the
Setpoint Timer Relay
Get the Status of the
Setpoint Thickness
Relay
COMPUTER INTERFACING
Page 5 - 10