Faults and troubleshooting
1.02
Laborota 4000/4001 efficient, 4010/4011 digital, 4002/4003 control
121
Fault/Message
Cause
Remedy
Main power switch turned off
Turn on main power switch
Main power switch defective
Contact service
Fuse defective
Lift is in end position
Use other arrow button
Mechanical or motor defect
Contact service
Motorised lift does not work
Vertical stop not correctly set
VAC Senso T / Rotavac vario
control power not switched on
Turn power switch on
No evacuation
Fuse defective
System not sealing properly
Check seals and connections
Check ground connections,
grease if required
Inadequate vacuum
Vacuum pump defective
Consult manufacturer's
documentation of vacuum
pump
Timer is programmed
Check or disable timer (see
chapter 6.5.2, page 116)
Instrument switches off
unexpectedly
(Laborota 4002 / 4003)
Fuse defective
Instrument switches off
unexpectedly
(Laborota 4002 / 4003),
Display shows SAFE
Emergency shutdown of
Laborota 4002 / 4003 control
when a pressure of 1100 mbar is
detected
Press any button to restore
the Laborota 4002 / 4003
control to its initial state
Calibrate vacuum sensor
(see chapter 8.1.1, page 122)
Table 7-1: Troubleshooting