beautypg.com

2 intended use and safety precautions 2 – Eppendorf InjectMan NI 2 User Manual

Page 8

background image

61

Before checking that the device has been delivered complete and installing it, please read the
following section thoroughly.

The control board must not be carried or dragged by the joystick!
The joystick is a highly sensitive control element which transmits movement commands in
the 0.1 µm range! Always carry the unit by the housing only!

Never place the control board upside down on the joystick

Treat the motor modules with care!

Never transport the motor modules while they are still assembled!

Whenever possible, transport the motor modules in their original packaging!

Do not tilt the assembled motor modules!

After the warm-up phase (at least half an hour) the device is virtually drift-free.

After switching on, wait for the initialisation process to complete before operating the joy-
stick. Completion of initialisation is indicated by an acoustic signal or by the LED returning
to its standby condition.

Enter an injection time >0.0 for the Eppendorf injection system being used, otherwise an
error message will be delivered

2.1 Hazards when unpacking and other information and instructions

Intended use and safety precautions

2 Intended use and safety precautions

2