– Error 044 – 046

In self-test wrong values defined for
displayed axis.

– Repeat self-test, reset, general reset.

If error recurs: Service

– Error 060 – 062

False data has been imported via the serial
ports.
There may be errors in the external PC
software.

– Transfer correct data.

– Error 098

Unable to write to EEPROM:
No internal data can be written to the
machine's EEPROM.

– General reset or Service (new

software download or hardware).

– Error 099

Internal Error: Internal error

– Reset, general reset.

If error recurs: Service