The following warnings or error messages may be displayed:

Remedy

– Warning 001

Check Injector and cable:
The injector is not ready or there is no
connection to the theoretically connected
injector.
Long injection path triggered at injector,
timeout of manipulator

– Make the injector ready or check the

connecting cable.

– Trigger "Inject" at manipulator with

long injection paths.

– Warning 002

Sync not immediate, limit or pressure:
An injection was triggered on the connected
injector even though the Move parameter is
entered on the InjectMan

®

NI 2 as the

injection synchronisation.

– Do not trigger the injection by the

injector in this case or change the
Synchon parameter.

– Warning 004

Connect motor module:
Occurs in self-test. The reason for this
warning is most likely that the motor module
is not connected to the control board.

– Connect motor module.

– Error 010 – 020

Certain hardware components of the control
board are missing.

– Service

– Error 021 – 023

Joystick hardware fault.

– Service

– Error 024

Wrong internal joystick parameters.

– Reset, general reset.

If error recurs: Service

– Error 025

For specific joystick axes wrong values are
defined. Reason may be an incorrect joystick
setting on power-up or in a general reset.

– General reset with joystick at mid

position.
If error recurs: Service

– Error 030 – 033

Internal software errors.

– Reset, general reset.

If error recurs: Service

– Error 034 – 039

Internal hardware faults.

– If recurring: Service

– Error 040 – 041

Internal software errors.

– Reset, general reset.

If error recurs: Service

– Error 042 – 043

Hardware faults

– If recurring: Service