5 user control and mode of operation, User control and mode of operation – Eppendorf InjectMan NI 2 User Manual
Page 41
![background image](/manuals/551253/41/background.png)
94
5.4.2.2 Inject
InjSpeed
: Speed of (axial) injection movement.
Adjustable from 7,500 to 5 µm/s
in 5-µm/s increments.
Synchron
: Synchronisation of connected
injector (e.g. FemtoJet
®
).
Setting options:
Immediate
: The injection pressure is
built upwhen the Inject function
is activated.
Limit
:
The injection pressure is built up when the (Z) limit is reached.
Move
:
No injection pressure is built up;
only the injection movement is executed.
Pressure
: The injection pressure is activated;
no injection movement occurs.
The synchronisation also applies to the special StepInj application.
Search+L
: When the parameter is active (ON), changing the limit value additionally changes
the position of the tool – and thus of the search plane – by the same value.
The distance between the limit and the search plane remains constant.
Setting options: ON / OFF
LimAxis
: The parameter sets the axis on which the limit is to be used.
Setting options: Z-axis / X-axis.
The Z axis setting should be chosen for adherent injection.
The X axis setting is used for special injections (not currently possible).
InjAxial
:
The active parameter (ON) causes an axial injection at the pre-set angle.
If the parameter is disabled, the injection is performed vertically downwards
(in Z direction with setting "LimAxis = Z-axis";
with "LimAxis = X-axis" the parameter is irrelevant).
Setting options: ON / OFF
--Submenu
Inject --
InjSpeed
7500µm/s
Synchron
LIMIT
Search+L
OFF
LimAxis
Z-AXIS
InjAxial
ON
▲
▲
5 User control and mode of operation
User control and mode of operation
5
05_Bedien_en.fm Seite 94 Donnerstag, 6. Dezember 2012 1:34 13