beautypg.com

07_fehler_en.fm, 7 troubleshooting 7 – Eppendorf InjectMan NI 2 User Manual

Page 47

background image

100

The following error situations are conceivable:

Remedy

– Motor axes move in wrong direction or do not

match joystick movement.

– Check installation parameters match

motor module setup, e.g. left side
mounting etc.

– Parameters for certain user actions are not

accessible.

– Enter parameters again or perform a

general reset and then reset the
machine.

– Machine not responding to key presses,

Home function still active.

– Press "Home" key again,

tool moves downwards
or move joystick.

– Tool does not move far enough down.

– Cancel limit function

or re-adjust tool

– Tool does not move straight downwards

when joystick is rotated, but also moves to
the side.

– Disable AXIAL function.

– Tool moves straight downwards during

injection

– Adapt InjAxial parameter.

– When connected, machine has no display

and will not switch on

– Check mains cable and plug.

If mains power supply unit defective:
Service

Troubleshooting

7 Troubleshooting

7

07_Fehler_en.fm Seite 100 Donnerstag, 6. Dezember 2012 1:37 13