Page 21

background image

74

According to the mounting mode on the left
side of the adapter.

Note

: In the "Install" menu (section 5.4,

subsection 5.4.2.6) the parameters
Side = LEFT
Direction = FLAT
Angle = e.g. 45°
are set in the given sequence according to the
mode of mounting.

4.3.2 Assembling the capillary holder

On the A module is the A head (1) for tool
mounting.

It is attached to the A module by two M3x16
screws as shown in the illustration.

The A head can accommodate tools with a
diameter of 4 mm, such as the universal
capillary holder (accessory FemtoJet

®

or

CellTram).

The universal capillary holder is inserted into
the A head after the knurled screw (arrow) has
been slackened, and then fixed in place by re-
tightening the knurled screw.

10

5

1

10

10

5

5

Installation

4 Installation

4

04_Install_en.fm Seite 74 Donnerstag, 6. Dezember 2012 1:32 13