Before cleaning the InjectMan
®
NI 2, disconnect the plug.
Ensure that no fluids enter into the inside of the micromanipulator in order to avoid short-circuits
in the electrical installation as well as corrosion.
Wipe painted parts and aluminum surfaces using a cloth and mild detergent and then with a dry
cloth.
Warning
:
Do not use any corrosive, solvent or abrasive detergents or polishes.
Clean the InjectMan
®
NI 2 before disinfecting.
Wipe every part of the InjectMan
®
NI 2, including accessories and connector cables, with an
alcohol-based disinfectant (Isopropanol, Alcohol).
Spray disinfection is not advisable, as disinfectants can enter into the micromanipulator.
The disinfection method used must comply with current disinfection regulations and guidelines.
If the device is to be inspected, repaired or calibrated by the service partners of the
Eppendorf AG, it must be free of hazardous substances and clean!
A decontamination certificate must accompany the InjectMan
®
NI 2.
A printed form ("Decontamination certificate for return of goods" and general instructions for
decontamination are available at our www.eppendorf.com homepage.
The serial number of the device must be entered in the decontamination certificate.
The dealer is responsible for warranty and service.
6.1 Cleaning
6.2 Disinfection
6.3 Decontamination prior to shipping
6.4 Service