75

To avoid having to reposition every time a capillary is changed, the
positioning aid can be mounted on the universal capillary holder so it fits
in the centre hole in the A head.

Note

: The connections between the control board, the mains power supply unit and the motor

module must only be plugged and unplugged with the device switched off.

Note

: Do not place the control board down resting on the joystick! When connecting cables,

place the control board on its side on an even surface and secure it against toppling over!

The connection to the motor module/injector (e.g. FemtoJet

®

) is made via the

corresponding connectors, as shown in Figure 4.

Check the voltage and frequency against the rating plate before connecting the mains
power!

Use only the mains power supply unit supplied by Eppendorf!

The cable is plugged into the mains power supply unit before being connected to the mains.

Important

:

The main power connection is the last one to be connected.

Check that the connectors are firmly attached.

Fig. 4: Underside of control board

1 Connection for optional foot switch (SER2; SubD9, male)
2 Connection for injector (FemtoJet

®

, Transjector 5246) (SER1; SubD9, male)

3 Connection for mains power supply unit

(

Important

:

Connect only after the other components)

4 Connection for motor module (SubD25, female)
5 Serial port (SER0; SubD9, male) for computer (RS 232)

4.4 Connecting cables – Startup

1

2

3

4

5

4 Installation

Installation

4

04_Install_en.fm Seite 75 Donnerstag, 6. Dezember 2012 1:32 13