4 installation, Installation – Eppendorf InjectMan NI 2 User Manual
Page 22
75
To avoid having to reposition every time a capillary is changed, the
positioning aid can be mounted on the universal capillary holder so it fits
in the centre hole in the A head.
Note
: The connections between the control board, the mains power supply unit and the motor
module must only be plugged and unplugged with the device switched off.
Note
: Do not place the control board down resting on the joystick! When connecting cables,
place the control board on its side on an even surface and secure it against toppling over!
–
The connection to the motor module/injector (e.g. FemtoJet
®
) is made via the
corresponding connectors, as shown in Figure 4.
–
Check the voltage and frequency against the rating plate before connecting the mains
power!
–
Use only the mains power supply unit supplied by Eppendorf!
–
The cable is plugged into the mains power supply unit before being connected to the mains.
Important
:
The main power connection is the last one to be connected.
–
Check that the connectors are firmly attached.
Fig. 4: Underside of control board
1 Connection for optional foot switch (SER2; SubD9, male)
2 Connection for injector (FemtoJet
®
, Transjector 5246) (SER1; SubD9, male)
3 Connection for mains power supply unit
(
Important
:
Connect only after the other components)
4 Connection for motor module (SubD25, female)
5 Serial port (SER0; SubD9, male) for computer (RS 232)
4.4 Connecting cables – Startup
1
2
3
4
5
4 Installation
Installation
4