Additional settings of density mode – RADWAG BALANCES 3Y SERIES User Manual
Page 93
![background image](/manuals/466046/93/background.png)
- 93 -
9. Load aluminium mass standard on balance’s weighing pan and on
stabilization of measurement result accept it by pressing <
> key,
10. The balance automatically calculates the density of air, and indicates its
value on the display,
11. Press <
> key to finish the procedure.
The value of determined air density is automatically assigned to field density> in menu weighing. 20.6. Additional settings of density mode The additional settings enable adjusting the working mode to user’s needs and requirements. To access the setting follow below procedure: Procedure: 1. Press grey coloured workspace, 2. The display indicates menu of: Settings, Keys, Information, Printout and Profile, 3. Press 4. The display indicates functions related to density determining process, such as: o Request sample number Available settings: NO – data on sample number is not required; applicable in case of a single measurement YES – before each measurement a window requesting inserting sample number is automatically displayed; applicable in case of a series of measurements o Measuring unit: the balance enables choosing between the following units: [g/cm 3 ], [kg/m 3 ], [g/l]. Selected unit is activated for all density determining options and also in final printout. o Assign density to a product: YES/NO – if option selected, the software automatically assigns the value of determined density of solid or liquid to field