Printouts – RADWAG BALANCES 3Y SERIES User Manual
Page 119

- 119 -
7. Minimum volume: for specifying the minimum volume of a pipette (in
case of a pipette with fixed volume, insert <0>);
8. Number of channels: for specifying number of pipette’s channels, in
case of a single channel pipette insert <1>;
9. Type: NONE/A/D1/D2. Pipette type is compatible with applicable
standard. Selecting pipette’s type is mandatory if calibration process
complies with the requirements of standard ISO 8655, as values of
errors differ in each type of a pipette. Therefore, if the software is to
accept adequate error values it is necessary to determine type of a
pipette;
10. Tested volumes: if pressed, a window is opened with a list of
calibrated volumes (in case of a new pipette, the list is empty), add
volume to be tested, and define error value for each tested volume:
•
Press <
Add > soft key
•
A window with on-screen numeric keyboard is displayed,
•
Insert value of tested volume in [µl] and accept by pressing <
> key,
•
A new record is automatically added to the list with recommended
error values,
•
In order to change error values press a field with added volume
of a pipette,
each field is editable and enables inserting user values,
Caution:
While in calibration process, the order of tested volumes corresponds to
the order in which they are entered.
•
On setting correct values return to the main window.
24.4. Printouts
Option Printouts enables setting the content of a standard printout as well
as the non-standard printout.
Standard printout
comprises four internal sections featuring different variables. For each
variable set option YES to include it in a standard printout. If option NO is
selected, then the variable is disabled, i.e. it is not printed.