Automatic adjustment time, Adjustment report printout, Glp project – RADWAG BALANCES 3Y SERIES User Manual
Page 54: Adjustment report printout 11.8. glp project
- 54 -
Time – adjustment takes place in time intervals declared in menu
Temperature – adjustment is triggered by temperature change only
Both – adjustment activation is triggered both by temperature changes
and time interval changes
CAUTION:
Changing the settings of automatic adjustment is enabled only in balances which are
not subject to conformity assessment (i.e. non-verified balances).
11.6. Automatic adjustment time
<
Automatic adjustment time > determines time interval in which
automatic internal adjustment of a balance is activated. The time interval is
declared in hours and ranges between 1 and 12 hours.
Setting time interval of automatic internal adjustment:
Select option < Automatic adjustment time >
Using displayed menu select appropriate time interval (given in hours)
which is a time gap elapsing from the last carried out internal automatic
adjustment until activating the following automatic internal adjustment.
CAUTION:
Changing the settings of automatic adjustment is enabled only in balances which are
not subject to conformity assessment (i.e. non-verified balances).
11.7. Adjustment report printout
Parameter <
Report printout > determines whether or not a report on
automatic internal adjustment should be automatically printed on its
completion.
Setting automatic report printout on adjustment completion. Go to parameter:
< Report printout > and select option
11.8. GLP project
GLP is one of means for maintaining documentation from work in accordance
with adopted quality system. Data selected for printing are printed on each
release of a report on balance adjustment.
Balance operator can use in a GLP report below listed information and signs:
adjustment (adjustment mode)
working mode (name of a working mode)
date, time,
user,
balance type,
balance ID,
level status,
nominal mass,
current mass
difference
temperature,