RADWAG BALANCES 3Y SERIES User Manual
Page 218

- 218 -
Accessible responses:
SM_OK CR LF
SM_I CR LF
- command carried out
- command understood but not accessible at this
moment (e.g.: other working mode, not PARTS
COUNTING)
ES CR LF
- command not recognized (incorrect mass format)
Set target mass value (e.g. for DOSING)
Format: TV_XXXXX CR LF, where: _ - space, XXXXX – mass format
Accessible responses:
TV_OK CR LF
TV_I CR LF
- command carried out
- command understood but not accessible at this
moment (e.g.: WEIGHING mode)
ES CR LF
- command not recognized (incorrect mass format)
Set reference mass value (e.g. for PERCENT SETUP)
Format: RM_XXXXX CR LF, where: _ - space, XXXXX – mass format
Accessible responses:
RM_OK CR LF
RM_I CR LF
- command carried out
- command understood but not accessible at this
moment (e.g.: other working mode, not PERCENT
SETUP)
ES CR LF
- command not recognized (incorrect mass format)
User log in
Format: LOGIN Name, Password CR LF (Name and password have to be
inserted as they are set in the balance – small and capital letters)
Accessible responses:
LOGIN OK CR LF
- command understood, new user is logged in
LOGIN ERRROR CR LF - command understood, an error in name or
password occurred, log in failed
ES CR LF
- command not recognized (error in format)
User log out
Format: LOGOUT CR LF
Accessible responses:
LOGOUT OK CR LF
- command understood, user is logged out
ES CR LF
- command not recognized (error in format)
Set balance mode
Format: PROFILE_Name CR LF where: _ - space (Name must be accordant
with the name form entered in a balance – lowercase and uppercase letters)
Accessible responses:
PROFILE OK CR LF
- command understood, new PROFILE activated