Proximity sensors
15.3.1.
The balance comprises two proximity sensors enabling wave control of the balance therefore
the user does not have to either press any key on the overlay or touch the screen.
The program recognizes 4 types of motions performed near the sensor:
1. Placing hand near the left sensor
2. Placing hand near the right sensor
3. Motion to the left
4. Motion to the right
CAUTION: using option LEFT MOTION or RIGHT MOTION, it is necessary to
remember that
ms!
To each of the motions, one of the available options may be assigned: [None; Profile;
Adjustment; Zeroing; Tarring; Tare setting; Tare switch off; Reset tare; Packaging; Print;
Header printing; Footer printing; Unit; Universal Variable 1; Universal Variable 2; Universal
Variable 3; Universal Variable 4; Universal Variable 5; Confirm; Abort; User; Left door; Right
door; Open/Close door; Parameters; Product, Warehouse; Client].
The selected option is recognized by the program and performed when activated by a
particular motion.
In order to ensure correct operation the user must remember to set the right sensors
sensitivity and the right value for proximity sensors delay (for more information refer
to point 33of this manual).
Automatically opened door
15.3.2.
This menu section comprises settings on automatic opening and closing
door of a balance featuring such mechanism, i.e. balances MYA 3Y and XA
3Y.A series. The settings may refer to the quick access keys as well as the
infrared proximity sensors located on the overlay of balance’s terminal.
Settings for a balance MYA 3Y series cover:
o
Left door / the option enables changing the position of the door on the left
side of the weighing chamber
o
Right door / the option enables changing the position of the door on the
left side of the weighing chamber
o
Open / Close door / the option enables changing the position of the door,
e.g. if the weighing chamber is opened, then activating the option causes
closing it. If the weighing chamber is closed, then activating the option
causes opening the door located on the right side of the weighing chamber,
Settings for a balance XA 3Y.A series cover:
o
Open / the option enables opening the door of the weighing chamber
o
Close / the option enables closing the door of the weighing chamber
o
Open / Close door / the option enables changing the position of the door,
e.g. if the weighing chamber is opened, then activating the option causes
closing it. If the weighing chamber is closed, then activating the option
causes opening the door of the weighing chamber.