RADWAG BALANCES 3Y SERIES User Manual
Page 121

- 121 -
the weighing chamber and the measuring vessel, thus stopping liquid
evaporation process.
Carried out research proved that the adapter for pipettes calibration eliminates
or considerably reduces liquid evaporation process, which is essential for correct
gravimetric calibration of piston pipettes.
Selecting an appropriate adapter and balance mostly depends on type of
calibrated pipettes.
Before starting pipettes calibration process, assembly the adapter inside
balance’s weighing chamber, and insert data on pipettes to be calibrated,
including complete pipette characteristics and values of errors (see: Database of
Pipettes).
The following step is setting process’s main option:
•
Number of measurements
•
Request batch number (YES/NO)
•
Operation with ISO 8655 (YES/NO)
•
Acquire ambient conditions from THB module (YES/NO)
Caution:
Description of options and settings are given in point 24.1 of this user manual.