Additional settings of pipettes calibration mode – RADWAG BALANCES 3Y SERIES User Manual
Page 117

- 117 -
1. Setup – access to balance’s menu
2. Print header
3. Print footer
4. Select pipette
The grey coloured workspace contains the following data:
o
Pipette
o
Tested volume
o
Average volume
o
Systematic error
o
Random error
o
Status
24.1. Additional settings of pipettes calibration mode
The additional settings enable adjusting the working mode to user’s needs and
requirements. To access the setting follow below procedure:
Procedure:
1. Press grey coloured workspace,
2. The display indicates menu of: Settings, Keys, Information, Printout
and Profile,
3. Press
4. The display indicates functions related to calibration of pipettes.
Pipettes calibration is related to the following functions:
−
Number of measurements; this option enables determining number of
measurements for tested volume of a pipette (valid for each tested
volume in case of a pipette with adjustable volume),
−
Request serial number: YES/NO. For option
calibration procedure, the software displays a window and on-screen
keyboard for inserting serial number of a pipette,
−
Comply with ISO 8655: YES/NO. For option
volumes, during pipettes calibration process the software automatically
accepts error values as specified in standard ISO 8655 (if other error
values have been accepted for a pipette, then on selecting option