Pipettes calibration – quick access keys, Adding a pipette to the database of pipettes – RADWAG BALANCES 3Y SERIES User Manual
Page 118
- 118 -
those from ISO standard),
−
Acquire data on ambient conditions from THB module: YES/NO. For
option
acquires and saves ambient conditions’ parameters from a connected THB
module. If option
the pipettes calibration process operator should manually insert ambient
conditions parameters (i.e. temperature, humidity and pressure in the
weighing room) from external ambient conditions sensors.
−
Result control, Tare mode, Automatic footer printout, Printout
mode, Printout, means of using these functions is given in point 10.8.
“Additional parameters on weighing process”.
24.2. Pipettes calibration – quick access keys
Each of working modes features a set of default quick access keys which are
automatically displayed on mode activation. The set of keys can be modified
by selecting other quick access keys to the bottom bar of the display. Such
process requires appropriate operator’s access level.
24.3. Adding a pipette to the Database of Pipettes
Database of pipettes contains list of pipettes’ names and other data including
tested volume and errors values for a specific volume. When adding a pipette
to the database, first specify its name, and then add other data on a pipette.
The software operates intuitively, and commands a user to follow displayed
information.
Adding a pipette to the Database can be carried out in menu of the database.
Procedure:
•
Enter submenu < Databases>, press <
Pipettes> field
•
Press <
Add > soft key, should a new pipette be added
The software automatically adds a new entry to the database, and enters its
editing mode. Insert all parameters of a new formulation.
List of parameters defined for a pipette:
1. Name: press “Name” field, a window opens for inserting name of a
pipette;
2. Code: for inserting pipette’s code;
3. Model: for inserting model name of a pipette;
4. Tip: for inserting name of used tip;
5. Type of volume: ADJUSTABLE / FIXED;
6. Nominal volume: for specifying nominal volume of a pipette;