RADWAG BALANCES 3Y SERIES User Manual
Page 219
- 219 -
LOGIN ERRROR CR LF
- command understood, an error in PROFILE name,
log in failed
ES CR LF
- command not recognized (error in format)
The result release
Format: SS CR LF
Accessible responses:
SS_OK CR LF - command understood, realization in progress
The command simulates pressing PRINT button on an overlay, according to
settings selected for result release.
Internal adjustment/calibration
Format: IC CR LF
Accessible responses:
IC_A CR LF
IC_D CR LF
- command understood, realization in progress
- calibration completed
IC_A CR LF
IC_E CR LF
- command understood, realization in progress
- time limit of result stabilization exceeded
IC_I CR LF
- command understood but not accessible at this
moment
Lock automatic internal adjustment
Format: IC1 CR LF
Accessible responses:
IC1_I CR LF
- command understood but not accessible at this
moment
IC1_E CR LF
- unavailable operation, e.g. verified balance
IC1_OK CR LF
- command completed
For verified balances the automatic internal adjustment cannot be locked.
For non-verified balances the command locks internal adjustment, it may be
unlocked either by IC0 command or by switching the balance off.
The command does not influence settings responsible for adjustment process
beginning.
Unlock automatic internal adjustment
Format: IC0 CR LF
Accessible responses:
IC0_I CR LF
- command understood but not accessible at this
moment
IC0_OK CR LF
- command completed
Function not available for verified balances.
Give balance factory number