beautypg.com

Input sensor correction constants (span & shft) 23, Setpoint limit values (splo & sphi) 23, Setpoint ramp rate (sprp) 23 – Red Lion TCU User Manual

Page 31

background image

Input Sensor Correction Constants (SPAN & SHFt)

If the controller temperature disagrees with a reference temperature

instrument or if the temperature sensor has a known calibration, the controller
temperature can be compensated by a correction slope (SPAN) and offset
(SHFt).

SPAN -

0.001 to 9.999

SHFt -

-999 to 9999

The following equation expresses the relationship:
Desired Display Temp = (Controller Temp x SPAN) + SHFt

Example 1:

The controller reads 293°F while a reference instrument

indicates 300°F. A SHFT value of +7°F corrects the controller indication
to match the reference. (Use SPAN = 1.000)

Example 2:

A thermocouple probe is calibrated over the region of operation

to achieve more accurate temperature control. The calibration results are as
follows:

Desired Temperature

Thermocouple Output

400.0°F

395.0°F

800.0°F

804.0°F

SPAN

F

F

F

F

=

° -

°

° -

°

=

800

400

804

395

0 978

.

SHFT = 400°F - (0.978 x 395°F) = 13.7°F

SPAN value of 0.978 and SHFT value of 13.7°F corrects the indicator to

the probe.

Setpoint Limit Values (SPLO & SPHI)

The controller has programmable high and low setpoint limit values to

restrict the setting range of the setpoint. Set the limit values so that the
temperature setpoint value cannot be set outside the safe operating area of the
process. On models equipped with Second Analog Input, configured as a
Remote Setpoint, the Remote Setpoint reading is also restricted to these
limits.

SPLO -

-999 to 9999

SPHI -

-999 to 9999

Setpoint Ramp Rate (SPrP)

The setpoint can be programmed to ramp independent of the controller’s

display resolution. The setpoint ramp rate can reduce thermal shock to the
process, reduce temperature overshoot on start-up or setpoint changes, or
ramp the process at a controlled rate.

SPrP -

0.1 to 999.9 degrees/minute

A ramp value of zero disables setpoint ramping. If the optional user input is

programmed for setpoint ramp, it affects the enabling and disabling of
setpoint ramping (See

User Input

, page 24). Setpoint ramping is initiated on

power-up or when the setpoint value is changed and is indicated by a decimal
point flashing in the far right corner of the main display.

Once the ramping setpoint reaches the target setpoint, the setpoint ramp

rate disengages until the setpoint is changed again. If the ramp value is
changed during ramping, the new ramp rate takes effect. If the setpoint is
ramping prior to invoking Auto-Tune, the ramping is suspended during
Auto-Tune and then resumed afterward using the current temperature as a
starting value. Deviation and band alarms are relative to the target setpoint,
not the ramping setpoint. If the analog output is programmed to transmit the
setpoint value, the instantaneous ramping setpoint value is transmitted.

Note: Depending on the ramp rate relative to the process dynamics, the actual

process temperature may not track the ramping setpoint value.

23

Figure 13, Setpoint Ramp Rate