DO Setup: Set Print Criteria
DO6000 and PCD6500 meters
Technology Made Easy ...
This screen allows you to select which criteria are printed with the measurement when you print
the data. The status of the current print criteria is displayed on the screen. The criteria option is
active if ON appears to the right of the option. It is inactive if OFF appears to the right of the
option. Any active criteria will be printed on demand.
To set Print Criteria
1. Touch the Touch here to edit button next to the Print Criteria to access the DO Print
Criteria Setup screen.
2. Touch ON or OFF to change the status of a criteria you want to modify.
3. Repeat step 2 for all the remaining criteria except Print Interval (or Print Option).
4. In the Print Interval (or Print Option):
Select TIMED to set the desired time interval in seconds via numeric keypad. Press
Enter to confirm time. The BS key allows you to remove characters that are incorrectly
entered. When TIMED is selected, data is sent to the printer buffer memory at the
selected time intervals and once the page is full, would be sent to the printer.
OR
Select LOG DATA to send all data stored in the data log memory to the printer when the
Print button in pushed from the measurement screen.
5. Touch the OK button to accept the changes of the entire group of print criteria and return
to the Setup screen.