Thermo Fisher Scientific CyberScan 6000 Series Meters v.1 User Manual
Page 51

39
pH Setup: View Stored Data
PH6000, PC6000, PH6500, PC6500 and PCD6500 meters
Technology Made Easy ...
The PH6000, PC6000, PH6500, PC6500 and PCD6500
have a memory capacity to store up to
1000 data points. The View Stored Data screen allows you to look at specific data points stored in
the meter based on the meters memory capacity.
The meter stores pH or pH (FET) data under the following parameters:
Reading mV
Measurement
Date / Time / Channel
Slope
Sample Id
Last Standardizations
Temperature
Meter Model / Serial No
Current Buffers
Operator
To View Stored Data
1. Touch the View button in the pH Setup screen.
Function buttons on View Stored Data Screen
Touch OK to go back to the pH Setup
screen from the View Stored Data
screen.
When you touch the Help button,
information about the current screen
appears. This information includes step-
by-step instructions for operating the
meter from the current screen and
possible applications information for that
screen.
Touch Delete to delete a selected data
point from the list. To delete a data point,
first touch the data point you want to
delete then touch the Delete button.
Touch Delete All to delete all the data
point in the memory.
Touch Print to print all the data points in
the memory.
See next page on Export View and
Header (or Export View) for
explanation.
You can only log 1000 data points at a time. To clear space for new data points, you
have to delete the same number of old data points as you want to add new data points.