Using the 6000 Series Meter: Button Functions
PH6000, PC6000, PH6500, CON6000, DO6000, PC6500 and PCD6500 meters
Technology Made Easy ...
The following button appears in all 6000 series meters single channel screens:
The Print button will send information to the output device that is connected to your
meter. If you selected the “Log data” option in the “Print criteria” setup, clicking on the
print button will dump the logged data to the printer.
The Standardize button accesses the standardization screen from the various
measure modes and initiates standardization of the meter once the standardization
screen is accessed. Not available from mV screen.
The Measure button directs the meter to measure your sample when in the Auto
Read function. Not available from mV, BOD, OUR, and SOUR screens.
The following buttons appear in the PH6000, PC6000, PH6500, PC6500 and PCD6500 meters
mV single channel screen:
The rel mV button returns the meter into the relative millivolt mode when in the
absolute mV mode. When this mode is activated, a window will appear asking you to
set a relative mV value or accept the default value of 0 mV. This feature may assist the
user to standardize certain analytical and monitoring activities such as titration.
The abs mV button returns the meter to absolute millivolt mode when the meter is in
relative mV mode. It only appears when the meter is in the relative mV mode.
The following buttons appear in the DO6000 and PCD6500 meters *Dissolved Oxygen
channel screen:
The BOD button allows you to enter the BOD Parameter Calculations Screen from the
BOD measurement screen.
The OUR button allows you to Start and Stop the OUR measurement from the OUR
measurement screen.
The SOUR button allows you to Start and Stop the SOUR measurement from the
SOUR measurement screen.
The following buttons appear in all 6000 series meters standardization screens:
The Confirm button accepts current value of the buffer or standard solution.
The Clear button clears all previous standardization points.
The Cancel button cancels current standardization and returns to the measurement
screen.
The Temp Std button allows you to check the accuracy of your temperature probe and
standardize to an accurate thermometer.