Thermo Fisher Scientific CyberScan 6000 Series Meters v.1 User Manual

Page 68

background image

56

Ion Setup: Set Default Temperature

PH6500, PC6500 and PCD6500 meters

Technology Made Easy ...

Set Default Temperature
It is a well known fact that ion concentration is a temperature dependent measurement. The
factory default setting is 25°C. If you are measuring a solution that is not 25°C and if you are not
using a temperature probe, then you should enter the temperature value of that solution in order
to get the correct concentration value. The current default temperature setting will be displayed
on the screen.

The default temperature can be set from -5°C to 105°C.

To set Default Temperature

1. Select temperature units by touching the appropriate unit button: C (Celsius), F

(Fahrenheit) or K (Kelvin).

2. Touch the Default Temperature box and use the numeric keypad to enter the desired

default temperature (-5°C to 105°C).

3. Press Enter in the keypad to return to Ion Setup screen.

Best results are obtained when the standards and the samples are at the same temperature.