Thermo Fisher Scientific CyberScan 6000 Series Meters v.1 User Manual
Page 16

4
6000 Series Meters Specifications
PH6000, PC6000, PH6500, CON6000, DO6000, PC6500 and PCD6500 meters
Technology Made Easy ...
cell constants range
0.1, 1.0, 10
conductivity
0 to 500 mS/cm over 4 ranges
0 to 200.0 µS, 200.0 µS to 2.000 mS,
2.000 to 20.00 mS, 20.00 to 500.0 mS
resistivity
3 ohm-cm to 100 megohm-cm over 4 ranges
0 to 20.00 KOhms, 20.00 to 200.0 KOhms,
200.0 KOhms to 2.000 MOhms,
2.000 to 100.0 Mohms
Total Dissolved Solids (TDS)
0.0 to 500.0 ppt over 4 ranges
(depending on the TDS factor)
0 to 200.0 x TDS factor ppm,
200.0 x TDS factor to 2000 x TDS factor ppm,
2.000 x TDS factor to 20.00 x TDS factor ppt,
20.00 x TDS factor to 200.0 x TDS factor ppt
salinity
0 to 90 ppt over 4 ranges
0 to 0.094 ppt, 0.094 to 1.000 ppt,
1.0
to 11.50 ppt, 11.50 to 90.00 ppt
accuracy
0.5% of full scale + 1 digit
(Applicable only for PC6000,CON6000, PC6500, and PCD6500)
range
0 to 60 mg/L (0 to 1272 mbar)
0 to 600 % saturation
resolution
0.01 mg/L (0.1 mbar)
0.1 % saturation
accuracy
0.1 + 1 least significant digit
(Applicable only for DO6000 and PCD6500)
range
-5.0 to +105.0 °C
,
(
Applicable for PC6000, PH6500, CON6000, PC6500, and PCD6500)
range
-5.0 to +46.0 °C
(Applicable for DO6000 & DO mode in PCD6500)
resolution
0.1
°C
accuracy
± 0.2 °C
range 450 to 825 mmHg
resolution
1 mm Hg
accuracy
±1 mm Hg + 1 least significant digit
(Applicable only for DO6000 and PCD6500)
range
0 to 45 ppt
(Applicable only for DO6000 and PCD6500)
operating temperature
5-45 °C
operation humidity
5-80 % noncondensing
maximum operating altitude
2000m
installation
category
II
Pollution category degree
2
(Applicable for PH6000, PC6000, PH6500, CON6000, DO6000, PC6500, and PCD6500)
Temperature Mode
Operating Conditions
Conductivity Mode
Dissolved Oxygen Mode
Barometric Pressure
Salinity Correction