pH Setup: Select Auto Read Mode

PH6000, PC6000, PH6500, PC6500 and PCD6500 meters

Technology Made Easy ...

Select Buffer Recognition
This option allows you to select Automatic buffer recognition or manual buffer recognition when
standardizing. With the automatic buffer recognition activated, the meter will automatically
recognize the buffers from the chosen buffer group and accept them when the meter recognizes
the reading as stable. When in the Manual buffer recognition mode, you must enter the buffer
value during the standardization procedure. The meter will accept the manually entered buffer
when you recognize that the measurement is stable. During the standardization procedure, you
may accept the buffer value before the meter recognizes it as stable by touching the
Standardization button.

To select Buffer Recognition

1. Touch MANUAL or AUTO to choose the method of buffer selection


Select Auto Read Mode
You can use this meter when the Auto Read function is active or when it is inactive. When the
Auto Read function is active, the meter will lock onto a reading when the meter recognizes it as
stable. The meter will not deviate from this reading until Measure is touched. If the Auto Read
function is inactive, then the meter will continuously monitor the pH of the sample and the
Measure screen display will indicate any fluctuation in the sample pH.

To select Auto Read Mode

1. Touch MANUAL or AUTO to choose the desired read mode.