beautypg.com

Thermo Fisher Scientific CyberScan 6000 Series Meters v.1 User Manual

Page 15

background image

3

6000 Series Meters Specifications

PH6000, PC6000, PH6500, CON6000, DO6000, PC6500 and PCD6500 meters

Technology Made Easy ...

640x480 digit LCD

screen size

4 1/2” x 6”

menu

options

extensive

help

screens

extensive

configurable

display

yes

keypad

controls

context specific touchscreen

(Applicable for PH6000, PC6000, PH6500, CON6000, DO6000, PC6500, and PCD6500)


inputs/outputs

BNC, ATC, DIN (FET)-

(PH6000,PC6000,PH6500,PC6500,PCD6500

)

DIN

(Conductivity)

– (

PC6000, CON6000, PC6500, and PCD6500)

DIN (DO)

-

DO6000, PCD6500,

RS232, USB-B, USB-A, RJ45 –

(Applicable for

all models)

electrical requirements

115 V/60 Hz, 230 V/50 Hz

output from PSU

9VDC, 3.3A center negative

line voltage tolerance

± 10%

input

impedance

>10

12

ohms

meter

size

6.75"x9.25"x3.5"

meter weight

2.8 lb.


1000 data pts

internal

diagnostics

yes

programmable data storage

store on stable, time, manual

programmable data output

timed or logged data

print interval

3 to 86400 sec

programmable

alarm

yes

(Applicable for PH6000, PC6000, PH6500, CON6000, DO6000, PC6500, and PCD6500

)

range

-2.000 to 20.000

resolution

0.1/0.01/0.001

relative accuracy

± 0.1/0.01/0.002 + 1 Lsd

automatic buffer recognition

yes

manual buffer recognition

yes

calibration

points

5

auto

lock

yes

FET

yes

(Applicable only for PH6000, PC6000, PH6500, PC6500, and PCD6500)


range

±

2000.0

resolution

0.1

accuracy

±

0.2

(Applicable only for PH6000, PC6000, PH6500, PC6500, and PCD6500)



range

1x10

–6

to 9.99x10

10

resolution

2 / 3 or 4 digits

relative

accuracy

±0.17n%

calibration

points

5

incremental methods

KA, KS, AA, AS

(Applicable only for PH6500, PC6500, and PCD6500)



continued on next page

Display

Memory

pH Mode

mV Mode

Ion Mode

General