beautypg.com

Thermo Fisher Scientific CyberScan 6000 Series Meters v.1 User Manual

Page 80

background image

68

DO Setup: Select Temperature Units / Select Barometer

DO6000 and PCD6500 meters

Technology Made Easy ...

Select Temperature Units
This setup function allows you to choose the temperature units show in the measurement screen
(celsius, fahrenheit or Kelvin).

To select Temperature Units

1. Select temperature units by touching the appropriate unit button: C (Celsius), F

(Fahrenheit) or K (Kelvin).


Set Barometer
The DO6000 and PCD6500

meters use an internal barometric pressure sensor in both the auto

and manual calibration mode. Prior to calibration check the actual barometric pressure. If this
value is different than the current barometric pressure displayed on the DO6000 and PCD6500,
you must calibrate the internal barometer. See below to set the barometer. DO NOT use the
barometric pressure issued by the weather service since this value is adjusted to sea level.

The barometer can be set from 450 to 825 mmHg or 599 to 1099 mbar.

To set Barometer

1. Touch mmHg or mbar to choose desired barometric units.
2. The current internal barometric pressure value is displayed on the Barometer box.
3. Touch the Barometer box and use the numeric keypad on the screen to enter the actual

barometric pressure. The BS key allows you to remove characters that are incorrectly
entered.

4. Touch Enter on the keypad to accept the barometric pressure value and return to the

Setup screen.