Thermo Fisher Scientific CyberScan 6000 Series Meters v.1 User Manual
Page 79

67
DO Setup: Select Stable Indicator / Select Auto Read Mode
DO6000 and PCD6500 meters
Technology Made Easy ...
Select Stable indicator
This setup option also lets you turn ON or OFF the stability indicator. When this function is turned
on, the STABLE appears when the DO reading is stable. This option also allows the user to
choose how fast (Fast, Med and Slow) the meter will accept a measurement or standardization
as stable. Choosing the Slow setting requires the meter to wait a longer time period before
accepting the measurement as stable, this reflects the highest level of precision.
To select Stable indicator
1. Touch ON or OFF to turn stable indicator on or off.
To select Stable indicator speed
1. Touch Fast, Med or Slow to select the stability indicator speed.
Select Auto Read Mode
This setup option allows you to select the activity of the Auto read mode. If the Auto read is active
(AUTO), the meter will freeze the reading on the screen when the measurement becomes stable.
To unfreeze the reading the Measure button must be pressed. If the Auto read is inactive
(MANUAL), the meter provides continuous monitoring of the sample. The stable indicator will
appear in manual read mode if the measurement is stable and the stable indicator is turned "ON"
but will not freeze the reading if there is any fluctuations in that reading.
To select Auto Read Mode
1. 1. Touch AUTO or MANUAL to choose the desired read mode.
NOTE: In BOD, OUR and SOUR setup modes, the Auto/Manual Read Mode is not
available. See BOD, OUR and SOUR Setup for details.