beautypg.com

KROHNE TIDALFLUX IFM 4110 PF EN User Manual

Page 55

background image

7.30830.32.00

55

Group S

Faults / Symptoms

Cause

Remedial action

Defective signaling instrument(s)
or external voltage source

Check signaling instrument(s) or
external voltage source and
replace if necessary.

Internal power supply (E+/E-) is
voltage source: shorted, one or
several pulse outputs defective.

Check connections and cables,
change if necessary (see sect.
2.5.6).
Voltage between E+ and E-
approx. 24 V.
Check fuse F8 on the I/O PCB
and replace if necessary (see
sect. 8.5). If instrument still
doesn't not operate, check fuses
F.. on the I/O PCB for the
status outputs and replace if
necessary:
F2 for terminals A1 and A

F3 for terminals A2 and A

F4 for terminals D1 and D

F5 for terminals D2 and D

If it still does not operate, one or
several pulse outputs are
defective.
Replace I/O PCB, see sect. 8.4.

Control inputs C1 and C2 are set
to “Hold outputs” or to “0”.

Change setting, see sect. 4.4
and 5.10, Fct. 1.11 and 1.12.

S1
(A1, A2, D1, D2)

No reaction from connected
signaling instrument(s)

In addition to this, the red LED
flashes = Fatal Error

Replace signal converter, see
sect. 8.3.

S2
(A1, A2, D1, D2)

Signaling instrument(s) is (are)
constantly tripped

“All Error” or “Fatal Error”
settings

Check settings in Fct. 1.07-1.10
and change if necessary, see
sect. 4.4 and 5.9.

Terminal “A1” not defined as
status output

Adjust in Fct. 3.07

S3
(only for A1)

No reaction of connected
signaling instrument

Wrong connection/polarity

Observe polarity for driver
capacity 0.1 < I

≤≤ 0.2 A

see sect. 6.3.
A1 = “+” and A

= “-“

S4
(only for A1)

Cyclic tripping of signaling
instrument

Terminal “A1” not defined as
status output

Adjust in Fct. 3.07.

Group C

Faults / Symptoms

Cause

Remedial action

Wrong connection

Connect properly, see sect.
2.5.5 and 2.5.6

Defective control input C or
voltage source (internal or
external)

Check connections and cables
and change or replace if
necessary. Check voltage
source. Check fuses F6 and F7
on I/O PCB and replace if
necessary.

C1

No function of control inputs

Wrong setting of control inputs

Change setting, see sect. 4.4
and 5.10.