KROHNE TIDALFLUX IFM 4110 PF EN User Manual
Page 30
7.30830.32.00
30
ADC-HARDW.
Hardware error A/D converter
Replace ADC printed circuit board
ADC GAIN
Hardware error A/D converter
Replace ADC printed circuit board
FC-HARDW.
Hardware error on field current PCB
Replace field current PCB
FATAL.ERROR
Fatal error, measurement interrupted
Replace electronic unit or consult
factory.
4.6
Resetting the totalizer and deleting error messages, RESET/QUIT menu
Delete error messages in RESET/QUIT menu
Key
Display
Description
- - - - - - -
- - - - - / - - -
Measuring mode
↵
CodE 2
- -
Key-in entry code 2 for RESET/QUIT menu:
↑
→
↑
→
ERROR QUIT.
Menu for error acknowledgement
→
QUIT. NO
Do not delete error messages,
press
↵
key twice to return to measuring mode
↑
QUIT. YES
Delete error messages
↵
ERROR QUIT.
Error messages deleted
↵
- - - - - - -
- - - - - / - - -
Return to measuring mode
Reset totalizer in RESET/QUIT menu
Key
Display
Description
- - - - - - -
- - - - - / - - -
Measuring mode
↵
CodE 2
- -
Key-in entry code 2 for RESET/QUIT menu:
↑
→
↑
→
ERROR QUIT.
Menu for error acknowledgement
↑
TOTAL.RESET
Menu for resetting totalizer
→
RESET NO
Do not reset totalizer,
press
↵
key twice to return to measuring mode
↑
RESET YES
Reset totalizer
↵
TOTAL.RESET
Totalizer is reset
↵
- - - - - - -
- - - - - / - - -
Return to measuring mode