KROHNE TIDALFLUX IFM 4110 PF EN User Manual

Page 41

background image

7.30830.32.00

41

Factors for volume F

M

Volumetric unit

TEXT VOL.

Factor F

M

Setting

Cubic meters

m

3

1.0

1.00000 E+0

Liter

Liter

1000

1.00000 E+3

Hectoliters

h Lit

10

1.00000 E+1

Deciliters

d Lit

10000

1.00000 E+4

Centiliters

c Lit

100000

1.00000 E+5

Milliliters

m Lit

1000000

1.00000 E+6

US gallons

USGal

264.172

2.64172 E+2

Millions US gallons

USMG

0.000264172

2.64172 E-4

Imperial gallons

GBGal

219.969

2.19969 E+2

Mega imperial gallons

GBMG

0.000219969

2.19969 E-4

Cubic feet

Feet3

35.3146

3.53146 E+1

Cubic inches

inch3

61024

6.10240 E+4

US barrels liquid

US BaL

8.36364

8.38364 E+0

US barrels ounces

US BaO

33813.5

3.38135 E+4

Factors for time F

T

Time unit

TEXT TIME

Factor F

T

Setting

Seconds

sec

1

1.00000 E+0

Minutes

min

60

6.00000 E+1

Hours

hr

3600

3.60000 E+3

Day

DAY

86400

8.64000 E+4

Year

YR

31536000

3.15360 E+7

5.15

F/R mode, forward/reverse flow measurement

Refer to Section 2.5.6 for electrical connection of outputs.

Define direction of forward flow, see Fct. 3.02, subfunction “FLOW DIR.”:

in conjunction with F/R operation, this is where to set the direction of the forward flow.
“+” means the same direction as shown by the arrow on the primary head.
“-“ means the opposite direction.

Set one of the status outputs to “SIGN I”, “SIGN P” or “SIGN P2”, see Fct. 1.08-1.10 (1.07).

Dynamic behavior of outputs in case of “SIGN I, P or P2” see Section 5.9.

Current and/or pulse outputs must be set to “2 DIR.”, see Fct. 1.05, 1.06 and 1.07,

subfunctions “FUNCT. I”, “FUNCT. P” and “FUNCT. P2”.