7.30830.32.00

49

7. Functional checks

7.1

Checking the zero with IFC 110 PF signal converter, Fct. 3.03

Set “zero” flow in the pipeline. Make sure that the measuring tube is completely filled with liquid.

Switch on the system and wait at least 15 minutes.

Press the following keys for zero measurement:

Key

Display

Description

If “YES” is selected in Fct. 3.04 ENTRY CODE
key in 9-stroke CODE 1 now:

.

Fct. 1.00

OPERATION

2x

Fct. 3.00

INSTALL.

Fct. 3.01

LANGUAGE

2x

Fct. 3.03

ZERO SET

CALIB. NO

CALIB. YES

0.00

- - - - - / - - -

Flow rate displayed in set unit, see Fct. 1.04 DISPLAY,
subfunction “DISP. FLOW”.
Zero is measured, duration approx. 15-90 s. “WARNING” is
displayed when flow is “>0”, acknowledge by pressing

key.

STORE NO

If new value is not to be stored, press

key (3 times)

(4 times = return to measuring mode).

STORE YES

Fct. 3.03

ZERO SET

Store new zero value

(2x) 3x

- - - - - - -

- - - - - / - - -

Measuring mode with new zero

7.2

Checking the measuring range Q, Fct. 2.01

For this test a measuring value can be simulated in the range of –110 to +110 percent of Q

100%

(set full scale range, see Fct. 1.01 FULL SCALE).

Switch on the system.

Press the following keys for checking the measuring range:

Key

Display

Description

If “YES” is selected in Fct. 4.04 ENTRY CODE
key in 9-stroke CODE 1 now:

.

Fct. 1.00

OPERATION

Fct. 2.00

TEST

Fct. 2.01

TEST Q

SURE NO

SURE YES

0

PCT.

Current, pulse and status outputs indicate corresponding values.

±

10

PCT.

±

50

PCT.

±

100

PCT.

±

110

PCT.

Fct. 2.01

TEST Q

End of test, actual measured values again available at outputs

(2x) 3x

- - - - - - -

- - - - - / - - -

Back to measuring mode

7.3

Hardware information and error status, Fct. 2.02

Before consulting the factory about errors or flow measurement problems, please invoke Fct. 2.02 HARDW. INFO

(hardware information).

An 8-character and a 10-character status code are stored under this function in each of 4 “windows”. These 8 status codes