KROHNE TIDALFLUX IFM 4110 PF EN User Manual
Page 19
7.30830.32.00
19
2.5.7
Standard factory settings
All operating data are set at the factory in accordance with the specifications contained in the order.
If no specifications are made in the order, instruments will be delivered with the standard parameters and functions indicated in
the table below.
To facilitate the start-up of the instrument, current and pulse outputs are set to handle measurements in “two flow directions” so
that the current flow rates and volumes are displayed and/or counted independent of the direction of the flow. The figures
displayed may have a preceding sign.
Such factory settings of current and pulse outputs may lead to measuring errors, particularly when volumes are metered and
totalized, for example if pumps are switched off and “backflows” occur which are not within the low-flow cutoff (SMU) range,
or if separate displays and counts are required for both flow directions.
To avoid faulty measurements, it may therefore be necessary to change the setting of the following functions:
-
SMU low-flow cutoff
Fct. 1.03.
-
display
Fct. 1.04.
-
current output I
Fct. 1.05.
-
pulse output P
Fct. 1.06.
Standard factory settings
Fct. No.
Function
Setting
Fct. No.
Function
Setting
1.01
Full-scale range
See instr. nameplate of
primary head
1.10
Status output D2
Indication F/R
1.02
Time constant
3 sec. for display,
pulse, current and
status outputs
1.11
Control input C1
Totalizer reset
1.03
Low-flow cutoff
OFF
1.12
Control input C2
OFF
1.04
Display
Flow rate
totalizer
m
3
/hr
m
3
3.02
Primary head
meter size
direction of flow
See instr. nameplate
+ direction, see arrow
on primary head
1.05
Current output I
function
range
error detection
2 directions
4-20 mA
22 mA
3.04
Entry code
NO
1.06
Pulse output P
Function
pulse value
pulse width
2 directions
1000 pulses/s
symmetric
3.05
User unit
Liter/hr
1.07
Pulse output 2, A1
function
pulse value
pulse width
2 directions
1 pulse/s
50 ms
3.06
Application
flow
ADC gain
special filter
pulsating
automatic
OFF
1.08
Status output A2
ON
3.07
Hardware
terminal A1
self check
pulse output A1
NO
1.09
Status output D1
All error