KROHNE TIDALFLUX IFM 4110 PF EN User Manual
Page 23
7.30830.32.00
23
4.3
Key functions
In the following, the cursor or flashing part of the display is shown against a gray background.
To start operator control
Measuring mode
operator control mode
PLEASE NOTE: if “YES” is selected in Fct. 3.04 ENTRY CODE, “CodE 1 ---------“
appears in the display after pressing the
→
key.
Enter the password for the entry code which is a sequence of 9 keys:
→→→↵↵↵↑↑↑
(each keystroke confirmed by “*”).
To terminate operator control
Press
↵
key any number of times until one of the following menus Fct. 1.0 OPERATION, Fct. 2.0 TEST or Fct. 3.0
INSTALL is displayed.
Press key
↵
Store new parameters: acknowledge by
pressing key
↵.
Measuring mode is continued with new
parameters.
New parameters not to be stored:
Press key
↑ to display “STORE NO”.
Measuring mode is continued with the “old”
parameters after pressing key
↵.
Keyboard with 10 keys
The keyboard with the 10 keys (0-9) is used for setting all flashing numbers (cursor).
Exception: the digits of the function numbers, such as Fct. 1.03, can only be changed with keys
↑
or
↓
.
To change numbers
Increase number
Decrease number
To shift cursor (flashing position)
Shift to right
1 3 . 5 7 1
m 3 / h r
F c t . 1. 0
O P E R A T I O N
F c t.
3 . 0
I N S T A L L .
S T O R E Y E S
3 9 7 . 3 5
m 3 / h r
3 9 7 . 4 5
m 3 / h r
3 9 7 . 3 5
m 3 / h r
3 9 7 . 3 5