beautypg.com

KROHNE TIDALFLUX IFM 4110 PF EN User Manual

Page 53

background image

7.30830.32.00

53

Group I

Faults / Symptoms

Cause

Remedial action

Display shows…
I SHORT
Current output shorted,
load < 15

Eliminate short circuit,
load must be

15

!

I OPEN
Load > 500

Find interruption and eliminate.

No information displayed after
test

I1

Receiver instrument indicates
“0”.
Invoke test function 2.03 for
analysis see Sect. 7.4

as described for faults I2 and I9

Wrong connection / polarity

Connect properly,
see sect. 2.5.2 and 2.5.6.

Defective circuit and/or receiver
instrument

Check circuit and receiver
instrument at I+ / I- and replace
if necessary.
Check fuse F9 on I/O PCB and
replace if necessary, see sect.
8.4 and 8.5.

Defective current output

Replace I/O PCB (see sect.
8.4) or consult Krohne Service,
having first noted down
hardware information and error
status, see sect. 7.3, Fct. 2.02.

Wrong flow direction setting

Set properly in Fct. 3.1.

I2

Receiver instrument indicates
“0”.

Current output switched off

Switch on in Fct. 1.5.

I3

22 mA are available at current
output (fault current)

Range of current output I is
exceeded

Check instrument parameters
and correct if necessary (see
sect. 2.5.2 and 5.7) or consult
Krohne Service, having first
noted down hardware
information and error status, see
sect. 7.3, Fct. 2.02.

I4

22 mA are available at current
output (fault current) and red
LED flashes

Fatal error

Replace signal converter or
consult Krohne Service, having
first noted down hardware
information and error status, see
sect. 7.3, Fct. 2.02.

I5

Unsteady display

Electric conductivity of process
liquid to low

Increase time constant (see
sect. 5.2, Fct.1.2). Also refer to
sect. 6.7.

I6

Receiver instrument indicates
“constant value”

Control input C1 or C2 is set to
“Hold outputs” and is activated

Change setting (see sect. 5.10,
Fct. 1.11 and 1.12), or
deactivate control input.

I7

Jumping current values

Current output is set to automatic
range change

Change hysteresis or tripping
ranges, see sect. 5.19.

I8

F/R mode:
different displays for identical
flow volumes in both directions

Different range set for “forward
flow” and “reverse flow”

Change setting, see sect. 5.15,
Fct. 1.05 “Rev. range”.

I9

Receiver instruments indicates
“min. values”.

Control input C1 or C2 is set to
“Zero outputs” or “Hold outputs”
and is activated

Change setting (see sect. 5.10,
Fct. 1.11 and 1.12) or deactivate
control input.