→ PULSWIDTH = select pulse width, press → key
•
AUTO
automatic = 50 % of cycle duration of 100 % output frequency
•
SYM.
symmetric = pulse duty factor 1:1 across entire range
•
SEC.
(variable) setting range 0.01 – 1.00 s.
Select with
↑ and ↓ keys.
Press
→ key to change to numerical setting.
Change flashing number (cursor) with keys
↑ and ↓. Use ← and → keys to shift cursor 1 place to left or right.
Flashing numbers can also be directly set with the 10-key keyboard.
Press
↵ key to change to subfunction “VALUE P” and/or “VALUE P2”.
→
→ VALUE P = set pulse value per unit volume
only appears when “PULSE/VOL.” is selected in “SELECT P”, press
→ key
•
XXXX PulS/m
3
•
XXXX PulS/liter
•
PulS/US.Gal
•
PulS/user-defined unit, factory setting = “Liter” or “US Mgal/day”, see sect.5.14.
Select with
↑ and ↓ keys.
Press
→ key to change to numerical setting, 1st number (cursor) flashes.
Set numerical value
•
XXXXsetting range depends on pulse width and full-scale range:
P
min
= F
min
/ Q
100%
P
max
= F
max
/ Q
Q100%
Change flashing number (cursor) with keys
↑ and ↓. Use ← and → keys to shift cursor 1 place to left or right.
Flashing numbers can also be directly set with the 10-key keyboard.
Press
↵ key to return to Fct. 1.06 PULS P or to Fct. 1.07 PULS2 A1.
or
→
→ VALUE P2 = set pulse value per unit time
only appears when “PULSE/TIME” is selected in “SELECT P”, press
→ key
•
XXXXPulSe/Sec
•
XXXXPulSe/min
•
XXXXPulSe/hr
•
XXXXPulSe /user-defined unit, factory setting = “hr” or “day”, see sect. 5.14
Select with
↑ and ↓ keys.
Press
→ key to change to numerical setting, 1st number (cursor) flashes.
Set numerical value
•
XXXXsetting range depends on pulse width.
Change flashing number (cursor) with keys
↑ and ↓. Use ← and → keys to shift cursor 1 place to left or right.
Flashing numbers can also be directly set with the 10-key keyboard.
Press
↵ key to return to Fct. 1.06 PULS P or to Fct. 1.07 PULS2 A1.
Please refer to section 2.5.7 “Standard factory settings”.
Refer to sect. 2.5.6 for connection diagrams and to sect. 5.16 for characteristics.