beautypg.com

KROHNE TIDALFLUX IFM 4110 PF EN User Manual

Page 36

background image

7.30830.32.00

36

→ PULSWIDTH = select pulse width, press → key

AUTO

automatic = 50 % of cycle duration of 100 % output frequency

SYM.

symmetric = pulse duty factor 1:1 across entire range

SEC.

(variable) setting range 0.01 – 1.00 s.

Select with

↑ and ↓ keys.

Press

→ key to change to numerical setting.

Change flashing number (cursor) with keys

↑ and ↓. Use ← and → keys to shift cursor 1 place to left or right.

Flashing numbers can also be directly set with the 10-key keyboard.

Press

↵ key to change to subfunction “VALUE P” and/or “VALUE P2”.

→ VALUE P = set pulse value per unit volume
only appears when “PULSE/VOL.” is selected in “SELECT P”, press

→ key

XXXX PulS/m

3

XXXX PulS/liter

PulS/US.Gal

PulS/user-defined unit, factory setting = “Liter” or “US Mgal/day”, see sect.5.14.

Select with

↑ and ↓ keys.

Press

→ key to change to numerical setting, 1st number (cursor) flashes.

Set numerical value

XXXXsetting range depends on pulse width and full-scale range:

P

min

= F

min

/ Q

100%

P

max

= F

max

/ Q

Q100%

Change flashing number (cursor) with keys

↑ and ↓. Use ← and → keys to shift cursor 1 place to left or right.

Flashing numbers can also be directly set with the 10-key keyboard.
Press

↵ key to return to Fct. 1.06 PULS P or to Fct. 1.07 PULS2 A1.

or

→ VALUE P2 = set pulse value per unit time
only appears when “PULSE/TIME” is selected in “SELECT P”, press

→ key

XXXXPulSe/Sec

XXXXPulSe/min

XXXXPulSe/hr

XXXXPulSe /user-defined unit, factory setting = “hr” or “day”, see sect. 5.14

Select with

↑ and ↓ keys.

Press

→ key to change to numerical setting, 1st number (cursor) flashes.

Set numerical value

XXXXsetting range depends on pulse width.

Change flashing number (cursor) with keys

↑ and ↓. Use ← and → keys to shift cursor 1 place to left or right.

Flashing numbers can also be directly set with the 10-key keyboard.
Press

↵ key to return to Fct. 1.06 PULS P or to Fct. 1.07 PULS2 A1.

Please refer to section 2.5.7 “Standard factory settings”.

Refer to sect. 2.5.6 for connection diagrams and to sect. 5.16 for characteristics.