9. Troubleshooting

74

9. Troubleshooting

Symptom Possible

Cause

Maintenance

Cannot obtain Zero Count

(When using a filter)

1) The inside of the optical system may

be dirty.

Attach the zero filter and perform purging for a

prolonged time.

After performing a measurement in a highly

concentrated environment, attach the zero filter and

perform purging.

2) The filter’s air tightness may be

deteriorated.

Replace your zero filter.

3) Leaks may be occurring inside the

instrument.

Return the instrument to your distributor or to your

KANOMAX service center for repair.

Count value is too high

(Higher than the expected)

1) The particle concentration at the

measurement site may be high.

-

2) The inside of the optical system may

be dirty.

Attach a zero filter and perform purging for a

prolonged time.

After performing a measurement in the highly

concentrated environment, attach the zero filter and

perform purging.

3) The instrument may be being used

outside the operating environment

defined in the specifications.

(Measurement Environment Temperature

/ Humidity / Concentration etc)

Use the instrument under the environmental

conditions set in the specifications.

If you have checked the above 1), 2)

and 3), and the instrument readings

are still too high, the instrument may

require calibration or repair.

Contact your distributor or your KANOMAX service

center for repair or calibration.

Count value is too low

(Lower than the expected)

1) The particle count at the measurement

site may be low.

-

2) The pump vacuum is low.

Pay attention to the pump performance.

Check if an error status is displayed on the screen.

3) The instrument may be being used

outside the operating environment

defined in the specifications.

(Measurement Environment Temperature

/ Humidity / Concentration etc)

Use the instrument under the environmental

conditions set in the specifications.

4) If you have checked the above 1), 2)

and 3), and the instrument readings

are still too low, the instrument may

require calibration or repair.

Contact your distributor or your KANOMAX service

center for repair or calibration..

The output of the option

sensors is not displayed.

1) The sensor may not be connected

properly.

Check the option sensor connection.

Touch panel becomes

unresponsive

1) The touch panel may be damaged.

Return the instrument to your distributor or to your