3. Display Description and Operation Procedure

16

3.5 MODE

Tap [MODE] to display the mode selection window.

*User needs to have the authorization
(Change Measurement Mode) to
change measurement mode.

In the Mode setting screen you can
configure measurement settings in each
mode.

Select a mode from below and tap it;

“SINGLE”,“CONTINUOUS”,“INTERVAL”,
“STATS”,“STANDARD”,”PRESET”,
“REMOTE”.

You can set items such as INTERVAL,
SAMPLE T., CYCLES, MANUAL STOP, AUTO
SAVE, AUTO PRINT and START DELAY.
Please note that the setting items vary
depending on the measurement mode.



3.5.1 SINGLE

[Measurement Flow]
Perform a measurement once within the pre-set time frame.
Example: If the setting is as follows:

INTERVAL: N/A
SAMPLE T.: 00:00:15
CYCLES: N/A

A 15 second sampling is performed.













* Delay Time: Set Delay Time in START DELAY screen (Refer to 3.8.1 ).

Sampling Time

(15sec)

Start

Measurement

Finish

Measurement

Wait Time

Pump ON

Delay Time

10 sec

Wait Time

*Note that the instrument is capable of 180 hours
of continuous operation in every measurement
mode.