beautypg.com

KANOMAX 3910 Portable Particle Counter User Manual User Manual

Page 43

background image

3. Display Description and Operation Procedure

35

3.6.3 STANDARD

- When not using a MAP file

Tap [START] to display “CONFIRM” screen.

A measurement will start based on the
settings configured in 3.5 MODE.









Tap [START] to start a measurement.

If “START DELAY” is NOT checked in

“MODE” setting:

Upon tapping [START], the pump will
begin running. After the pump has run for
10 seconds to stabilize, the measurement
will begin.

If only “START DELAY” is checked in

“MODE” setting:

Upon tapping [START], the Delay Time
begins. During the last 10 seconds of that
Delay Time the pump will run to stabilize,
after which the measurement will begin.

If “START DELAY” and “Set time”

are checked in “MODE” setting:

Upon tapping [START], “Input START
TIME” dialog box will be displayed. Enter
the time when you want to start
measuring and tap [OK]. During the last
10 seconds of the [Time left] time the
pump will run to stabilize, after which the
measurement will begin.


Tap [CANCEL] to stop measuring and go
back to the menu screen.


If you fail to tap [OK] after entering
“START TIME”, and the entered START
TIME arrives, the instrument will increase
the start time in 10 minute intervals until
you press [OK].