KANOMAX 3910 Portable Particle Counter User Manual User Manual
Page 30

3. Display Description and Operation Procedure
22
3.5.4 STATS
[Measurement Flow]
Configure SAMPLE T., CYCLES and SAMPLE POINT, and repeat measuring.
In STATS mode, INTERVAL and SAMPLE T. are configured with the same value. Measurements are
repeated according to the CYCLES settings in each SAMPLE POINT.
As for the measurement condition, you can configure the settings.
[Measurement Settings]
If there are any MAP files saved, they are
displayed in the dropdown list of SCHEDULE.
When measuring by using a MAP file, select a FILE
from the list.
Tap the entry field for “SAMPLE T”. to display a
numeric keypad. Enter “hours”, “minutes” and
“seconds” on the numeric keypad and tap [OK].
Tap the entry field of “CYCLES” to display a
numeric keypad. Enter a value and tap [ENTER].
Once a MAP file is loaded, you cannot change the
settings for SAMPLE POINT. (The configured
number of measurement points in the MAP file
will be applied.)
When a MAP file is not loaded, tap an entry field of
“SAMPLE POINT” to display a numeric keypad.
Enter a value and tap [ENTER].
[PRESET]
Tap “PRESET” to display a window asking you to
input “PRESET NAME”.
Tap the entry field to display the keypad. Enter
the file name and tap [Enter].
Tap [OK] to save the current settings on the MMC
card (Authentication required).
* In the future you can automatically load the
configuration saved here as a PRESET. Refer to
3.5.6 PRESET.
Check “AUTO SAVE” to store the data
automatically.
Check “AUTO PRINT” to print the data
automatically.
Check “START DELAY” to use a delay time
prior to a measurement.
Check “Set time” to set the time when you
want to start a measurement.
Tap [OK] to save and activate the settings
(Authentication required).