3. Display Description and Operation Procedure

39

Select a Location and tap [DETAIL] to view
the image of the selected Location. (The
image of the Location needs to be registered
when creating a Schedule using the provided
PC application.)


Tap [START] to start a measurement.

If “START DELAY” is NOT checked in

“MODE” setting:

Upon tapping [START], the pump will begin
running. After the pump has run for 10
seconds to stabilize, the measurement will
begin.

If only “START DELAY” is checked in

“MODE” setting:

Upon tapping [START], the Delay Time
begins. During the last 10 seconds of that
Delay Time the pump will run to stabilize,
after which the measurement will begin.

If “START DELAY” and “Set time” are

checked in “MODE” setting:

Upon tapping [START], “Input START TIME”
dialog box will be displayed. Enter the time
when you want to start measuring and tap
[OK]. During the last 10 seconds of the
[Time left] time the pump will run to
stabilize, after which the measurement will
begin.


Tap [CANCEL] to go back to the menu
screen.


If you fail to press [OK] after entering
“START TIME”, and the entered START TIME
arrives, the instrument will increase the
start time in 10 minute intervals until you
press [OK].


After closing the measurement results, tap

button while measuring to display a

chart.