3. Display Description and Operation Procedure

33

Tap [START] on the MAP to start a
measurement

If “START DELAY” is NOT checked in

“MODE” setting:

Upon tapping [START], the pump will
begin running. After the pump has run for
10 seconds to stabilize, the measurement
will begin.


If only “START DELAY” is checked in

“MODE” setting:

Upon tapping [START], the Delay Time
begins. During the last 10 seconds of that
Delay Time the pump will run to stabilize,
after which the measurement will begin.

If “START DELAY” and “SET time”

are checked in “MODE” setting:

Upon tapping [START], “Input START
TIME” dialog box will be displayed. Enter
the time when you want to start
measuring and tap [OK]. During the last
10 seconds of the [Time left] time the
pump will run to stabilize, after which the
measurement will begin.

Tap [CANCEL] to stop measuring and go
back to the menu screen.

If you fail to press [OK] after entering
“START TIME”, and the entered START
TIME arrives, the instrument will increase
the start time in 10 minute intervals until
you press [OK].

Tap button while measuring to
display a chart.

Tap [STOP] to stop a measurement.