KANOMAX 3910 Portable Particle Counter User Manual User Manual
Page 15

3. Display Description and Operation Procedure
7
3.2 Log-In
3.2.1 Log-In Display
Type your user ID and password in the
field and press [OK] button.
Press [Shut Down] button to turn off
the instrument.
[CAPS] button: To type upper case letters
[SWITCH] button: Jump to next field
[CLR] button: Clear input
[BS] button: Back space, delete the preceding character
3.2.2 User
The instrument manages the usage log based on user ID, manage users from the user setting
display (Refer to User Management).
User information consists of User name, User ID, Password, and Qualification.
User name: a user name with up to 40 characters
User ID: a user ID with up to 16 characters.
Password: The instrument recognizes users with the combination of user ID and
Password. Passwords need to be 4 to 16 alphanumeric characters. Passwords
expire every 180 days, after which the user ID will be invalid until a new
password is set.
Qualification: Each user is granted authorization to perform certain operations.
For example, if a user has authorization for [Change Measurement Setting]
operation, the user cannot change the measurement settings. If the current
logged in user does not have the authorization to carry out a certain operation,
the button for that operation is deactivated. Even if a user has authorization,
the user will need to enter his/her user name and password every time before
initiating an action that requires authorization.
The administrator can reset passwords for any users in case of a lost or forgotten password (Refer
to 3.8.3. Other - User Management).
If the administrator forgets his/her password, nobody can operate the instrument. Kanomax
strongly recommend registering multiple administrator users on the instrument.