3. Display Description and Operation Procedure

44

3.6.5 Saving Data


If AUTO SAVE is not checked on the MODE setting window, data will not be saved to a file after
measuring. If you try to change the MODE or start another measurement when AUTO SAVE is not
checked, you will get the message shown below.




- Tap [YES] to save the data.
- Tap [NO] to discard the data.







If the electricity goes out during a measurement, data will be treated as below.

[SINGLE] and [CONTINUOUS] Mode

Data during a measurement will be discarded when the power goes OFF.

[INTERVAL] Mode

Data taken before electricity goes out remains stored in the instrument.
When restarting the instrument, the message shown below will be displayed.


- Tap [YES] to save the data.
- The calculation data will be created using the

data collected before the power went out.

- Tap [NO] to discard the data.








[STATS] and [STANDARD] Mode

Data taken before electricity went out remains in the instrument. However, the data from the
location where a measurement was being performed when the power went out is not saved.
The measurement will be in a temporary state of suspension.

-> Refer to 3.6.6 In Case that Measurement is Interrupted.