14-29
IM DR232-01E
Specifications
14
14.10 Communication Interface Module
Type
Description
GP-IB
RS-232-C
RS-422-A/RS-485
Ethernet
Type
DT300-11
DT300-21
DT300-31
DT300-41
General Specifications
Power consumption
Included in the main unit or the subunit to which the module is to be installed.
Normal operating temperature/humidity
20 to 80%RH for 0 to 40
°C, 10 to 50%RH for 40 to 50°C(no condensation)
GP-IB Module
Style Number: S4
Electrical and mechanical specifications
conform to IEEE Standard 488-1978
Code
ISO (ASCII) code
Address
0 to 15
Functions
· Talker functions
Output of measurement values (ASCII, binary); output of setting parameters (ASCII)
· Listener functions
Setting of measurement conditions, control sof starting and stopping measurement, specifying
causes of interrupts (excluding setting and control of power ON/OFF)
Causes of interrupts
Syntax error, chart-end, completion of A/D conversion, operations of internal timer, end of data
saving/reading, incomplete measurement during computation
External dimensions
Approx. 57 (W)
× 137 (H) × 54 (D) mm
Weight
0.3kg
RS-232-C Module
Style Number: S8
Electrical and mechanical specifications
conform to standard EIA RS-232-C
Connection method
point-to-point
Communication method
half-duplex
Synchronization mode
Start-stop synchronization (synchronized by a start and a stop bit)
Baud rate
150, 300, 600, 1200, 2400, 4800, 9600, 19200 bps
Start bit
Fixed at 1 bit
Data length
7 or 8 bits, selectable
Parity
Selectable from even, odd, or none
Stop bit
1 or 2, selectable
Transmission distance
max. 15 m
Connector
D-sub 25pin