Channel -10, Set ok – Yokogawa DR240 User Manual
Page 113
5-10
IM DR232-01E
5.4 Configuring the Measuring Range and
Recording Span or the Linear Scaling of a
Strain Input Channel
D I S P
M O D E
P R I N T
F U N C
I N S
D E L
R A N G E A L A R M
S E T
C H A R T
E N T E R
F E E D
M. F U N C 1
M. F U N C 2
RECORD
ALARM
CHART
KEYLOCK
R E C O R D
001-01:STRAIN/200k
> Select Channel Number
SPAN=–20000/ 20000 * 10
µε
E S C
• Press the RANGE key to enter the RANGE menu.
• Using
and
, select and/or enter a value for each of the shaded fields shown below.
• To exit any of the following menu items during the procedure, press the MODE (ESC) key.
This returns to the first item of the menu. Note, however, that your new settings and selections
are canceled.
• Reaching the step showing the message “***SET OK***” confirms the latest settings/
selections.
Setting the Measuring Range and Recording Span
001-01:STRAIN/200k
>Select Channel No.
001-01:STRAIN/200k
>Select Channel No.
(Main menu item)
001-01:STRAIN/200k
001-01:STRAIN/200k
2k 20k 200k
SPAN=–20000/_20000
>Span limit(–20000 to 20000*10
µε
)
***SET OK***
(configuration complete)
ENTER
ENTER
ESC
Setting Linear Scaling
001-01:STRAIN/200k
>Select Channel No.
001-01:STRAIN/200k
>Select Channel No.
001-01:STRAIN/200k
(Main menu item)
(Submenu items)
001-01:SCL:STRAIN/200k
001-01:SCL:STRAIN/200k
2k 20k 200k
SPAN=–20000/_20000
>Span limit(–20000 to 20000 *10
µε
)
SCL=-100.00/_100.00
***SET OK***
(configuration complete)
ENTER
ESC
ENTER
(Press twice)
ENTER