Yokogawa DR240 User Manual
Page 238
13-5
IM DR232-01E
Trouble-shooting
and Maintenance
13
13.4 Error Codes
104
Illegal character(s) or code(s) is used before
Set the equation correctly.
or after an operator.
105
The right and left spans are the same.
Change the span.
106
Attempted to make range setting during
Stop making range setting.
computation.
110
Syntax error for computation constant
Set the constant correctly.
111
Out of constant setting range
Set the constant within the range.
120
Floppy disk not inserted or not formatted.
Insert a formatted floppy disk into the
drive.
121
Insufficient memory
Delete unnecessary files.
122
Incorrect file name or volume name
Set the correct file name or volume name.
123
Attempted to make settings which are not
Stop making the settings.
allowed to be made during saving or reading
of measured data.
124
Too many files in the media
Delete unnecessary files.
125
Attempted to read data which was saved
Select data which has been saved
by another instrument model.
by DR/DA.
127
Attempted to save data in a write-protected file.
Cancel write-protection.
128
No event/action has been set.
Set the desired action .
130
Incorrect channel found during saving
Set the correct channel.
or reading of measured data.
131
RAM disk error or floppy disk is ejected while
Clear the error and attemt the same operation
bein accessed.
or contact your nearest sales represantative.
132
Attempted to format the disk while a file is open.
Close the file.
134
Set-up data saving error
135
Internal data error
136
Incorrect copy channel
Set the correct copy channel.
137
Computation is started while set-up data is
Stop Computation.
being saved or read.
138
Floppy disk error
Use another floppy disk.
139
Other media related error
140
Sub-units selected which are not connected,
Specify correct sub-units.
or the power is not supplied.
141
Disconnected modules were selected.
Specify correct modules.
143
Modules selected which cannot be calibrated.
Specify correct modules.
144
Calibration done improperly.
Check electrical connections.
Calibrate again.
If the same error occurs again, contact
your nearest Sales representative.
145
Hardware becomes faulty during calibration.
Turn off the power and then on.
Calibrate again.
If the same error occurs again, contact
your nearest Sales representative.
170
The IP address does not belong to any of the
Set the correct IP address.
classes, A, B, or C.
171
There is a hole in the mask or the host
Set the correct subnet mask.
address section is not released.
172
The net address section including the subnet
Set the correct net address.
does not match the subnet section of the IP
address.
173
The host address section of the IP address is
Set the correct IP address.
either all zeroes or all ones.
ROM ERROR
System ROM error
Contct your nearest sales
representative.
ROM ERROR*
System ROM error
Contct your nearest sales
representative.
RAM ERROR
Main memory error
Contct your nearest sales
representative.
RAM-DISK ERROR
RAM disk error
Contct your nearest sales
representative.
FLOPPY ERROR
Floppy disk error
Contct your nearest sales
representative.
MAIN NV READ ERROR
Internal nonvolatile memory read error
Contct your nearest sales
representative.
MAIN NV WRITE ERROR
Internal nonvolatile memory write error
Contct your nearest sales
representative.
CARRIAGE CANNOT MOVE
Printer error
Contct your nearest sales
representative.
SYSTEM ERROR
System error
(
: number)
Check whether each unit or
(
: number)
power code is connected
properly after turning off the
power switch. If the same
error occurs again even though
you turn on the power switch,
contact your nearest sales
representative.
RESET ERROR
An error occurs between the main unit
Same as abeve.
and communication module.